25
Artis zee
Floor-mounted system for surgical angiography
Injectors
MEDRAD Avanta*
Contrast medium syringe 150 ml
Contrast flow mode fixed 1 to 45 ml/s, increment 1 ml/s; variable 1-10 ml/s in 0.1 increments
Fixed rate flow for saline flushing 1.25 ml/s
Adjustable rise time 0.1 to 9.9 seconds, 0.1 s increments
Contrast syringe refill user selectable 25 to 150 ml, increments of 25 ml, refill rate 2.75 ml/s
Automatic or manual refill 2.5 ml/s
Pressure range 20.5 to 82.7 bar
Syringe 150 ml
Feedback on actual injection parameters
Mechanical construction movable stand, removable injector head
Rack mount version
Injector head on overhead tube support, swivel only or swivel/movable mount
MEDRAD MARK V ProVis*
Contrast medium syringe 150 ml
Flow rates for 150 ml syringes 0.3 – 10.0 ml/s, ml/min/hr in 0.1 ml/s increments
10 – 50 ml/s in 1 ml/s increments
0.3 – 10.0 ml/min/hr in 0.1 ml/min/hr increments
10 – 59 ml/min/hr in 1 ml/min/hr increments
Release delay for injection or radiation 0 to 99.9 s in 0.1 s increments
Pressure limit 6 to 82 bar, corresponds to 100 to 1200 psi
Cylinder 150 ml
Double head injector, feedback on actual injection parameters
Mechanical construction movable stand, removable injector head
Rack mount version
Ceiling-mounted injector head, swivel only or swivel/movable mount
* Option