10
COMMAND LIST GAIN
COMMAND SYNTAX LIST G
ARGUMENTS None
DESCRIPTION Lists the 16 temperature gains. These are used to convert temperature
counts to degrees Celsius. This is the “M” term in the characterization
equation:
EXAMPLE To view the temperature gain settings:
Type: LIST G<CR>
The DSA will return the temperature gain settings. They could appear as
follows.
SET TEMPM0 793.
SET TEMPM1 432.
SET TEMPM2 441.
SET TEMPM3 429.
SET TEMPM4 402.
SET TEMPM5 400.
SET TEMPM6 413.
SET TEMPM7 400.
SET TEMPM8 410.
SET TEMPM9 412.
SET TEMPM10 421.
SET TEMPM11 430.
SET TEMPM12 430.
SET TEMPM13 412.
SET TEMPM14 422.
SET TEMPM15 432.
NOTE The temperature gain settings are determined when the sensor is
manufactured. They normally will not change after a calibration
COMMAND LIST IDENTIFICATION
COMMAND SYNTAX LIST I
ARGUMENTS None
DESCRIPTION Lists the IDENTIFICATION configuration variables.
EXAMPLE To verify the general module configuration settings:
Type: LIST i<CR>
The DSA will return:
SET ECHO 1
NOTE A user must be very careful when modifying one of these variables. An
incorrect value in one of these variables could have a detrimental affect on
the operation of the module.