DSA3200 Boot Parameter Modification
The DSA3200 Series modules use an operating system licensed from VxW orks. The boot parameters are
set at the factory, but these parameters may be modified to suit the needs of a specific installation. This
section documents the modification of the boot parameters.
1. De-energize the DSA3200. Connect the DSA3200 trigger/serial test cable (Scanco part
#155829) from the DSA3200 to a host PC.
2. Start HyperTerminal, or an equivalent communication program. Connect the DSA3200 to
a COM port on the PC Host. Set the serial parameters to 9600 BAUD, no parity, 8 data bits
and 1 stop bit.
3. Energize the DSA3200. The following sign-on information should be displayed.
VxWorks System Boot
Copyright 1984-1997 Wind River Systems, Inc.
CPU: HITACHI hs7709
Version: 5.3.1
BSP version: 1.1/0
Creation date: Feb 9 2000, 12:41:05
Scanivalve (c)2000, Boot loader version 1.00
Press any key to stop auto-boot...
4. Press any key within 3 seconds to stop the auto-boot process. If a key is not pressed within
the 3 seconds time, the system will proceed to auto-boot with the existing operating system
using the current setup.
NOTE: The boot-loader will continue to reboot until stopped by pressing a key under the
following conditions:
A. If boot-from-flash is selected, and no operating system is present
in flash.
B. if boot-from-net is selected and the FTP server is not correctly set
up.
5. W hen the auto-boot process is stopped, the boot-loader will prompt with:
[VxW orks Boot]:
To get a list of the existing boot parameters:
Type: p <Enter>
The following list is the default setup:
boot device : cs
processor number : 0
host name : host
file name : c:/dsa_hs/vxworks.st
inet on ethernet (e) : 191.30.80.100
host inet (h) : 191.30.101.109
user (u) : DSA_HS
ftp password (pw) : scanivalve
flags (f) : 0x0
other (o) : flash,000.096.093.218.000.002,10baset